AC Milligauss Meter Model
UHS2
AC Milligauss Meter Model UHS2 ---- $350
(Worldwide Free Shipping) 3-Axis
/ 1-Axis ELF VLF magnetic field meter
A digital gaussmeter for precise measurement of AC magnetic field.
Details:
This meter measures AC magnetic field in a wide frequency range
of 13 Hz to 75 kHz (75,000 Hz). The frequencies include most ELF (that is, frequencies below 1 kHz) and VLF (above
1 kHz) magnetic fields. The meter measures the true 3-axis magnitude of the AC field. Two other more specialized
measurements can also be performed: 3-axis VLF-only (1kHz – 75 kHz) to measure the strength of higher-frequency
fields, and 1-axis full bandwidth (13 Hz – 75 kHz) to determine the principal direction of the AC magnetic field.
These specialized measurements are selected by the knob labeled "TYPE OF MEASUREMENT". A "RANGE" knob allows a wide
range of field strengths to be measured—from .01 to 1999 milligauss.
Most measurements will be performed with both knobs turned to
the left-most position. This setting is for measurement of true magnitude of magnetic field at the widest range of
frequencies (ELF + VLF) up to a strength of 19.99 milligauss. If the field becomes stronger than 19.99, most of the
display will go blank except for a "1" at the extreme left. This indicates over-range. Then switch the left-side
switch to 199.9 (or 1999 if the 199.9 setting still shows over-range).
Quick Start
Instructions:
Flip the power switch (on right side) upward. Turn the two knobs to the left-most position (19.99 and 3-AXIS ELF
+ VLF). This will read the 3-axis magnitude of the magnetic field in milligauss (RMS equivalent). Avoid rapidly
rotating or rapidly tipping the meter; this causes some additional signal to be picked up from the earth's magnetic
field. The additional signal will be gone as soon as you stop moving the meter. This effect will be obvious once
you start using the meter. If the display shows a number "1" on the left side with no other numbers, it is
over-range. Then switch the left knob to the 199.9 or 1999 range.
If the right-side knob is set to 3-AXIS VLF, the meter will generally read much lower, or zero; this setting
reads only the higher-frequency magnetic fields (1,000Hz - 75,000 Hz), and not the full bandwidth (13 Hz - 75,000
Hz) that is read in the "3-AXIS ELF + VLF" setting. The third position of the right-side knob is 1-axis only, but
at full bandwidth (13 Hz - 75,000 Hz). It measures magnetic field only in one axis: the direction that corresponds
to the thin axis of the meter.
When the battery becomes low,
an additional decimal point (on the left side) will be displayed. Slide off the back door of the meter to replace
the standard 9-Volt battery.
Full Length Instructions:
This meter is designed to do precise measurements of AC magnetic field in a wide frequency range of 13 Hz
to 75 kHz (75,000 Hz). The frequencies include most ELF (that is, frequencies below 1 kHz) and VLF (above 1
kHz) magnetic fields. The meter measures the true 3-axis magnitude of the AC field. Two other more
specialized measurements can also be performed: 3-axis VLF-only (1kHz – 75 kHz) to measure the strength of
higher-frequency fields, and 1-axis full bandwidth (13 Hz – 75 kHz) to determine the principal direction of the AC
magnetic field. These specialized measurements are selected by the knob labeled “TYPE OF MEASUREMENT”. A
“RANGE” knob allows a wide range of field strengths to be measured—from .01 to 1999 milligauss.
Most measurements will be performed with both knobs turned to the left-most position. This setting is for
measurement of true magnitude of magnetic field at the widest range of frequencies (ELF + VLF) up to a strength of
19.99 milligauss. If the field becomes stronger than 19.99, most of the display will go blank except for a “1” at
the extreme left. This indicates over-range. Then switch the left-side switch to 199.9 (or 1999 if the 199.9
setting still shows over-range).
To use the meter, slide the power switch (on right edge) upward. With both knobs at their left-most
position, the display can read up to 19.99 milligauss. You’ll notice that if you suddenly rotate the meter or
tap it, the display will read higher for a fraction of a second. This is caused by the Earth magnetic field,
which is about 500 milligauss DC. Normally a DC (non-oscillating) field will not be detected by this meter;
however, any motion (tipping the meter or tapping it) will modulate this DC field, creating a temporarily-changing
field, which will be detected. The effect is especially noticeable in this very sensitive 19.99 range. To
avoid this false signal, keep the meter body stable and always oriented the same way in space as you walk along
making measurements. For example, if the meter is vertical and the slide switch on the side is pointing, let’s
say, northwest, then you should keep the meter always vertical with the power switch side pointing northwest, as
you walk along. While rotating or changing orientation of the meter, the reading will usually be higher than
it should be, but the reading will return to the correct value as soon as you stop rotating it. The meter will
also read a higher level if it's being moved near magnetized metal.
If the “TYPE OF MEASUREMENT” switch is set to its center position, the meter will read 3-axis VLF fields (at
frequencies of 1 kHz and higher). Generally in video displays and fluorescent lights a large percentage of the
fields are at these higher frequencies. Please note three factors when measuring these higher frequencies: 1) These
VLF measurements should be less than (or at most the same as) the measurement obtained when the knob is set on its
left-most position of “3-AXIS ELF + VLF”. In most areas, the VLF (with knob set in center) will be very low or
zero; at most it will never be more than the reading obtained when the RANGE knob is switched to “3-AXIS ELF +
VLF”. In other words, VLF alone is never more than ELF + VLF. 2) The center position of that knob is slightly
sensitive to 60 Hz fields, even though it is supposed to reject frequencies below 1 kHz. The rejection is 400:1.
That is, if there is a field of 400 milligauss that is exclusively at 60 Hz, the VLF (center) position of the knob
will yield a (false) reading of 1.0 milligauss. Similarly a 200 milligauss field at 60 Hz will yield a false VLF
field of 0.5 milligauss. 3) If the reading on ELF + VLF (left position of knob) is too high, the VLF reading may
not be correct when the RANGE knob is set at a lower number for the VLF reading. Specifically, if the ELF + VLF
(left side) reading is more than 40 milligauss, do not do a reading of the VLF-only (center position) with the
RANGE switched back to 19.99. Instead, use the 199.9 RANGE. Similarly, if the ELF + VLF is higher than 400
milligauss, continue to use the 1999 RANGE when you switch to VLF. Otherwise, the VLF reading may be falsely
low.
When the “TYPE OF MEASUREMENT” switch is set to the 1-AXIS position (right side), the meter reads ELF +
VLF but in only one axis. The direction of this 1-AXIS is the same as the “thickness” dimension of the meter
(as opposed to the left-right “width” or up-down “height” that are the conventional definitions of width and
height, with respect to the writing on the front label). This 1-axis sensor is just behind the center of the
black panel that’s on the top surface, or “roof” of the meter. The 1-AXIS measurement can be used to determine
the direction in which the field is strongest; unfortunately, this does not exactly correspond to the direction (or
the position relative to the meter) of the field source. The direction of strongest field can give you some
information about the direction of the field source, as per the convention developed by Edward Leeper.
For your reference, the other two magnetic sensors are 1) on the extreme right side of the meter’s “roof” (This
is the position of the sensor in the up-down direction.), and 2) the extreme left side under the “roof” (The
position of the left-right sensor). Generally, if you are very close to a field source, these three sensors will be
at three different distances from the field source, and as a result, the reading will be strongly dependent on the
angle at which you hold the meter. For field measurements more than about two feet from the source, this angle
dependence is minimal.
Accuracy: This meter has a typical error of +/-3% of the reading in the
frequency range 45 Hz to 5 kHz. At certain measurement angles and certain frequencies in the 3-AXIS modes, the
error may be as high as +/-7%. That is, a reading of 10.00 milligauss means the field is as high as 10.70 or as low
as 9.30, but most of the time the actual field will be between 10.30 and 9.70 (this is in RMS units). Note that
this accuracy specification is for sine waves in the very wide frequency range of 45 Hz to 5 kHz. Other
manufacturers of digital 3-axis meters only specify the accuracy for 60 Hz sine waves. (Those types of meter will
generally have significant additional error above 100 Hz; typically there is 5-10% additional error due to change
in sensitivity at other frequencies, making a total error that exceeds 10% above 100 Hz. Because of advanced
high-frequency detection circuitry, the model UHS does not have this problem). If you're measuring a signal that is
simultaneously stronger than 100 milligauss and higher frequency than 10 KHz, some additional inaccuracy may occur.
(It may read either too high or too low). An unusual generator with both a high frequency and high power output is
required in order to produce this type of magnetic field. You'll know that this is a problem if, as you approach
the strong field source, it eventually exceeds 1000 milligauss both on ELF + VLF and on VLF. The meter works by
measuring the absolute value of the AC field strength in 3 axes (or 1 axis), and then finding the time-average of
that absolute value. That time-average is then scaled so that a sine wave reads in RMS units. Sensitivity is down
to ½ at the frequency limits 13 Hz and 75 kHz. In the 3-AXIS 1 kHz - 75 kHz (VLF) setting, maximum error is +/-7%
of the reading at 5 kHz, and half sensitivity is 1 kHz and 75 kHz.
|